Product Details:
|
Product Name: | ASTMF2101 Mask Bacterial Filtration Efficiency BFE Detector Of Medical Face Mask Materials | Aerosol Chamber Specifications: | 600 X 85 X 3mm(L*W*H) |
---|---|---|---|
Cabinet Negative Pressure: | (-50~-200)Pa | Standards: | ASTMF2101, EN14683-2014 |
Operating Temperature: | (0~50)℃ | Warranty: | 1 Year |
Highlight: | fire test chamber,ultrasonic spot welding machine |
Technical index:
Main parameters |
Parameter range |
Resolution |
Maximum allowable error |
A channel sampling flow rate |
28.3L/min |
0.1L/min |
better than ±2.5% |
B channel sampling flow rate |
28.3L/min |
0.1L/min |
better than ±2.5% |
Spray flow |
(8~10)L/min |
0.1L/min |
better than ±5.0% |
Peristaltic pump flow |
(0.006~3.0)mL/min |
0.001ml/min |
better than ±2.5% |
Front pressure of channel A |
(-20~0)kPa |
0.01kPa |
better than ±2.5% |
Front pressure of channel B |
(-20~0)kPa |
0.01kPa |
better than ±2.5% |
Pressure before spray flowmeter |
(0~300)kPa |
0.1kPa |
better than ±2.5% |
Aerosol chamber negative pressure |
(-90~-120)Pa |
0.1Pa |
better than ±2.0% |
Operating temperature |
(0~50)℃ |
||
Cabinet negative pressure |
(-50~-200)Pa |
||
Data storage capacity |
>100000 groups |
||
HEPA filter characteristics |
Filtration efficiency for particles above 0.3um ≥99.99% |
||
Aerosol generator mass median diameter |
Average particle diameter (3.0 ± 0.3) μm, geometric standard deviation ≤1.5 |
||
Dual 6-stage Anderson sampler captures particle size |
Ⅰ>7μm,Ⅱ(4.7~7)μm,Ⅲ(3.3~4.7)μm,Ⅳ(2.1~3.3)μm,Ⅴ(1.1~2.1)μm,Ⅵ(0.6~1.1)μm |
||
Aerosol chamber specifications |
(Length 600 x diameter 85 x thickness 3) mm |
||
Total number of positive QC sampler particles |
(2200±500)cfu |
||
Negative pressure cabinet ventilation flow |
≥5m3/min |
Contact Person: Sophia Su
Tel: +86-13266221899